Gradient Formation:
Single quaternary or dual ternary
low pressure gradient
Flow Rate Range:
1–10,000 µL/min isocratic;
up to 6000 µL/min gradient
Flow Rate Precision:
<0.1% RSD at 1 mL/min
System Pressure Range:
0.1–35 MPa (5000 psi)
pH Range:
1–13
Sample Capacity:
3x well-plate-sized sample containers and
15 × 10 mL vials for reagents, diluents,
and transport liquids
Injection Volume:
0.1–250 µL, depending on loop and
syringe
Injection Methods:
Full-loop, partial-loop, microliter pickup,
and user-defned injection procedures
Fractionation Modes:
Time slices, peak signal, manual trigger,
external trigger
Fractionation Containers:
96 and 384 (deep) well plates
(un)capped vials, 0.3-10.0 mL
Delay Volume:
15 µL
Sample Thermostatting:
From 4 to 45 °C
Column Compartment:
Foced-air design with up to two
biocompatible 10-port switching valves
Column Thermostatting:
From 5 to 70 °C
UV Detection:
Variable wavelength or diode array
detector with biocompatible analytical or
semi-micro fow cells
Wetted Parts:
Titanium, sapphire, ruby, PEEK, PTFE,
UHMW polyethylene, Tefzel
®, Tefon® AF,
PPS, zirconium oxide, aluminum oxide,
PCTFE
Software Control:
Chromeleon Chromatography Data
System provides full control and trace-
ability for all components and automated
diagnostic tests
GLP Features:
System wellness monitoring and record-
ing of all system parameters in audit
trail by Chromeleon software; Tracking
system for columns, lamps, and fow
cells; Automated instrument qualifca-
tion (AutoQ™) with qualifcation period
monitoring