Light source: | DNP Standard color Viewer lamp (Four 10W dedicated fluorescent tubes are used) | Power supply: | AC100V 50, 60Hz | Lighting method: | High Frequency lighting method 20KHz | Color rendering: | Over 90% rendering evaluation-point for each color | Color temperature: | (Measured by Minolta CL-200) | Operating temperature: | 10~40℃ | Operating humidity: | Below 80%(avoid dew) | Major diameter dimension: | 375W×365H×165D m/m | Luminance plane dimension: | 250W×190H m/m | Color chart dimension: | 320W×290H m/m | Power consumption: | 80W | Weight: | 6.5kg/7.5kg | | | | CCV-31F CCV-51F | | | CCV-31F/G CCV-51F/G | | | |
|
●The Luminance plane characteristics: according to the types of self-contained fluorescent lamp |
|
Product Type | Luminance Color Temperature | Lamp | Color Temperature | Luminance | Color matching shift | Weight | CCV-31F | Fixed | FL10WW-DLB | 3100K±300 | Over 1300cd/m2 | Within 10% | 6.5kg | CCV-51F | Fixed | FL10D-EDL-56K | 5100K±300 | Over 1400cd/m2 | Within 10% | 6.5kg | CCV-31F/G | Volume adjustable | FL-10WW-DLB | 3100K±300 | Over 850~1300cd/m2 | Within 10% | 7.5kg | CCV-51F/G | Volume adjustable | FL10D-EDL-56K | 5100K±300 | Over 850~1400cd/m2 | Within 10% | 7.5kg | |
|
*The color temperature if Luminance. Color Temperature Conversion Type viewer ranges over 50k according to the change of the Luminance. *The specification may change without notice |
|
|
|
【1】 | Before using the product: | | | Power supply | | AC 100V, 50/60Hz shared | Light source | | 10W DNP Standard color Viewer Lamp (4) | Power consumption | | 80W | |
|
【2】 | Using the DNP Standard color Viewer | | | (1) | Plug the power cable of the product into AC100V | | | 【Precautions】 | Do not connect any power supply other than AC100V If the other power supply connected by accident, the inverter inside will burned out | | | (2) | Turn on the power switch | | (3) | Lamp lights: the lamp light will stabilised after 15 to 20 minutes at the room temperature(20℃) | | (4) | If the lamp does not lit, call for Service Office for the confirmation as described on other documents. | | (5) | To suspend snaking of the lamp, press power switch 3 to 4 times [as ON/OFF]. | |
|
【3】 | If the light does not lit | | | Call for Service Office to confirm the following: | | (1) | AC 100V supplied to the wall socket. | | (2) | Plug the product into the wall socket. | | (3) | Turn the power supply switch of the product ON | | | | | | | Still the lamp does not lit: Consider the following: | | | | | | | (1) | Two of the viewer lamps lit simultaneously, one of the fuses of the inverters has blowouts. | | (2) | Four of the viewer lamps does not light means that each of the fuses of the two inverters has blowouts. | | | 【Precautions】 | The above occurs if the applied power supply is not AC 100V. Fuse bum out cause internal burnout of inverter. The burned inverter needs to replace. | | |
|
【4】 | How to replace the lamp and keep it clean | | | Place the Viewer on the flat table, and plug out the power supply cable from the wall socket. Remove the used device in the following order: | | (1) | Unfasten the two of grub screws on the top frame. | | (2) | To remove the top frame from the main unit; Holding 50 to 60 mm the grub screw of the top frame and press it forward. | | (3) | Remove the glass plate and acrylic plate. | | | 【Precaution】 | Glass plate easily breaks and may hurt your hands and fingers. Handle the edge of the plate with care not to scratch. | | | | ◆Replacing the lamps: | | (1) | When the lamps used over 2,000 hours or one year, it needs to replace. | | (2) | Remove the Lamp Tubes: Hold the edge of the Lamp Tubes, and turn it 90 degree. | | | ◆Cleaning the device: | | (1) | Impregnate alcohol to gauze then apply it onto the surface of the reflection plate / acrylic plate / glass surface to remove the dirt. | | (2) | Clean the device once in six months to one year to keep quality of the Color Temperature and Luminance. | |
|
【5】 | Other Precautions | | | Handling the viewer needs the following precautions because it contains the parts using glass. | | (1) | Do not add strong impact on the viewer | | (2) | Do not put any heavy items on the surface of the viewer | |